RF303 Electrosurgical Analyzer Operators Manual PN 2202027 April 2007 © 2007 Fluke Corporation, All rights reserved. Printed in USA All product name

Pagina 3 - Technical Support

1-1 Chapter 1 Introduction and Specifications Contents Page Introduction ...

Pagina 4 - Manufacturing Location

RF303 Operators Manual 1-2

Pagina 5 - Table of Contents

Introduction and Specifications Introduction 1 1-3 Introduction The RF303 Electrosurgical Analyzer, hereafter referred to as the “Analyzer”, tes

Pagina 6

RF303 Operators Manual 1-4 Symbols Table 1-1 describes the symbols used in this document. Table 1-1. Symbols Symbol Description Symbol Description W

Pagina 7 - List of Tables

Introduction and Specifications General Safety Information 1 1-5 • Do not plug in or remove a test lead from either the ESU or the Analyzer wh

Pagina 8

RF303 Operators Manual 1-6 costs. The limits established by the following organizations are designed to provide reasonable protection against harmful

Pagina 9 - List of Figures

Introduction and Specifications Control and Interface Panels 1 1-7 Control and Interface Panels Figure 1-1 and Table 1-2 describe the controls

Pagina 10

RF303 Operators Manual 1-8 Table 1-2 Controls and Connections (cont.) Item Name / Description 4 Test Load (OHMS) Lamp Displays resistance in ohms. 5

Pagina 11 - Chapter 1

Introduction and Specifications Instrument Specifications 1 1-9 Instrument Specifications Modes of Operation Line Powered (Battery Charge and Ma

Pagina 12

Warranty and Product Support Fluke Biomedical warrants this instrument against defects in materials and work-manship for one full year from the date o

Pagina 13 - General Safety Information

RF303 Operators Manual 1-10 Test Load Section Main Test Load Selections...15 Selection range ...

Pagina 14 - W X Warning

Introduction and Specifications General Specifications 1 1-11 Front-Panel Controls / Pushbuttons Measurement Select (1) Load Select: Increment t

Pagina 15

RF303 Operators Manual 1-12 Power Requirements Universal input switching supply (12 V dc output) Operating Voltages: Specified...

Pagina 16

Introduction and Specifications Accessories 1 1-13 Accessories Table 1-3 lists the accessories provided with the Analyzer and their correspondin

Pagina 17 - Control and Interface Panels

RF303 Operators Manual 1-14

Pagina 18

2-1 Chapter 2 Installation, Setup, and Maintenance Contents Page Installation and Setup ...

Pagina 19 - Instrument Specifications

RF303 Operators Manual 2-2

Pagina 20

Installation, Setup, and Maintenance Installation and Setup 2 2-3 Installation and Setup The Analyzer has a universal power supply and automati

Pagina 21 - General Specifications

RF303 Operators Manual 2-4 exz02.eps Note The letter E, followed by a number indicates an error condition; the Analyzer has encountered a problem d

Pagina 22

Installation, Setup, and Maintenance Installation and Setup 2 2-5 Operating the Analyzer Using Battery Power You can operate the Analyzer using

Pagina 23 - Accessories

Unpacking and Inspection Follow standard receiving practices upon receipt of the instrument. Check the shipping carton for damage. If damage is found,

Pagina 24

RF303 Operators Manual 2-6 Table 2-2. Battery Operating Status Analyzer Connected to ac Power Source? On/Off Switch Setting Power Source to Analyzer I

Pagina 25 - Chapter 2

Installation, Setup, and Maintenance Installation and Setup 2 2-7 LED Backlight Display Operation The Analyzer’s display lights when you select

Pagina 26

RF303 Operators Manual 2-8 W X Warning To avoid possible electric shock, burning of the skin, or personal injury, follow these guidelines: Retractable

Pagina 27 - Installation and Setup

Installation, Setup, and Maintenance Installation and Setup 2 2-9 Note The mechanical pin is removed from the CQM connector to disable CQM duri

Pagina 28

RF303 Operators Manual 2-10 Maintenance Follow the recommendations below to keep the Analyzer in good working condition. Cleaning Moisten a clean clot

Pagina 29

Installation, Setup, and Maintenance Maintenance 2 2-11 exz16.eps Figure 2-1. Fuse Cover W X Warning To avoid severe electrical shock, disconne

Pagina 30 - Battery Status Light

RF303 Operators Manual 2-12 5. Reinstall the fuse carrier into the power inlet module, ensuring that the arrows on the ends of the fuse carriers poin

Pagina 31 - Analyzer

3-1 Chapter 3 Operation Contents Page The Signal Averaging Mode (SAM)... 3-3 Testing the ESU...

Pagina 33 - ESU CQM Safety Lead

Operation The Signal Averaging Mode (SAM) 3 3-3 The Signal Averaging Mode (SAM) Upon power-up, the Analyzer defaults to the instantaneous algor

Pagina 34 - Maintenance

Returns for partial refund/credit: Every product returned for refund/credit must be accompanied by a Return Material Authorization (RMA) number, obta

Pagina 35

RF303 Operators Manual 3-4 Note The Analyzer can test a wide range of electrosurgical units for basic operation and performance. It is compatible with

Pagina 36

Operation Testing the ESU 3 3-5 2. Connect the ESU Neutral/Dispersive electrode to the left blue jack dispersive, using the test lead, Part # 2

Pagina 37 - Operation

RF303 Operators Manual 3-6 To measure the ESU’s power output: 1. Activate the ESU. 2. If the WATTS indicator lamp on the panel is lit, go to Step 3.

Pagina 38

Operation Testing the ESU 3 3-7 SupplyMainsDispersiveActiveMainTestLoadESU RF303ESU ActiveHFMeterESU DispersiveEarthYELBLUBLUMCQM exz06.eps Figu

Pagina 39 - Testing the ESU

RF303 Operators Manual 3-8 Test Procedures The procedures involved with this test include: • connecting the ESU neutral/dispersive electrode to the A

Pagina 40 - ● Analyzer panel light is on

Operation Testing the ESU 3 3-9 To observe the ESU’s visual or audible alarm: 1. With the 50-ohm test load attached to the ESU CQM (REM) input,

Pagina 41

RF303 Operators Manual 3-10 HF Leakage Current Test 1 The HF leakage current test number one is an IEC isolated output/bipolar, CF-type leakage curren

Pagina 42

Operation Testing the ESU 3 3-11 3. If operating under battery power, attach a test lead from the set (Part # 1903307) from the battery ground

Pagina 43

RF303 Operators Manual 3-12 W Caution To avoid damage to the Analyzer and the ESU under test, do not exceed the Analyzer’s duty cycle, which is 30 sec

Pagina 44 - Test Procedures

Operation Testing the ESU 3 3-13 HF Leakage Current Test 2 The HF leakage test number two is an IEC earth-referenced, BF-type leakage current te

Pagina 45 - Figure 3-2. Basic CQM Check

i Table of Contents 1 Introduction and Specifications... 1-1 Introduction ...

Pagina 46 - HF Leakage Current Test 1

RF303 Operators Manual 3-14 WATTSmAOHMSENTMODE SELECTOHMS SELECT exz07.eps To measure the ESU’s signal HF leakage: 1. If the mA indicator lamp on the

Pagina 47

Operation Testing the ESU 3 3-15 SUPPLYMAINSEARTHDISPERSIVEESUDispersiveACTIVEGROUNDEARTHREFERENCEHFMETERESUEarth Reference Leakage Type BF Test

Pagina 48

RF303 Operators Manual 3-16 HF Leakage Current Test 3 The HF leakage test number three is an IEC earth-referenced, BF-type leakage current test. You p

Pagina 49 - HF Leakage Current Test 2

Operation Testing the ESU 3 3-17 WATTSmAOHMSENTMODE SELECTOHMS SELECT exz07.eps To measure the ESU’s signal HF leakage: 1. If the mA indicator

Pagina 50

RF303 Operators Manual 3-18 SUPPLYMAINSEARTHDISPERSIVEESUDispersiveACTIVEMAINTESTLOADGROUNDEARTHREFERENCEHFMETERESUEarth Reference Leakage Type BF Tes

Pagina 51 - 3-15

Operation Connecting an Oscilloscope to the Analyzer 3 3-19 Connecting an Oscilloscope to the Analyzer You can connect an oscilloscope to the A

Pagina 52 - HF Leakage Current Test 3

RF303 Operators Manual 3-20 Setting Up and Operating the Analyzer in Simplex Mode In Simplex Mode, you control operation of the Analyzer from the pane

Pagina 53

Operation Remote Operation via the RS-232 Function 3 3-21 2. Set the computer to communicate with a baud rate of 2400 with 8 data bits, no par

Pagina 54

RF303 Operators Manual 3-22 • All responses are uppercase, followed by CR and LF. • Measurement readings from the Analyzer are returned identically

Pagina 55

Operation Remote Operation via the RS-232 Function 3 3-23 Table 3-2. Available Duplex Mode Commands Command Description Parameters Required Ret

Pagina 56

RF303 Operators Manual ii HF Leakage Current Test 2 ... 3-13 HF Leakage Current Test 3

Pagina 57

RF303 Operators Manual 3-24 Table 3-2. Available Duplex Mode Commands (cont.) Command Description Parameters Required Return Value (Response) RDMODE

Pagina 58

Operation Remote Operation via the RS-232 Function 3 3-25 Table 3-2. Available Duplex Mode Commands (cont.) Command Description Parameters Requ

Pagina 59 - 3-23

RF303 Operators Manual 3-26 Table 3-2. Available Duplex Mode Commands (cont.) Command Description Parameters Required Return Value (Response) Prompt

Pagina 60

Appendices Appendix Title Page A Load Issues ... A-1 B Interpreta

Pagina 62

A-1 Appendix A Load IssuesLoad Issues The load resistors typically used in ESU analyzers are not ideal; they have some reactive components that are

Pagina 63 - Appendices

RF303 Operators Manual A-2

Pagina 64

B-1 Appendix B Interpretation of Fluctuating ReadingsInterpretation of Fluctuating Readings The RF303 Analyzer incorporates an accurate digital meas

Pagina 65 - Load Issues

RF303 Operators Manual B-2

Pagina 66

C-1 Appendix C AbbreviationsAbbreviations The following list includes abbreviations used in this document. A ampere ANSI American National Standar

Pagina 67 - Readings

iii List of Tables Table Title Page 1-1. Symbols ... 1

Pagina 68

RF303 Operators Manual C-2 in inch K kilo (103) kg kilogram kHz kilohertz ke kilohm lb pound LED light-emitting diode LCD liquid crystal display M m

Pagina 69 - Abbreviations

RF303 Operators Manual iv

Pagina 70

v List of Figures Figure Title Page 1-1. Analyzer Control and Interface Panels ............ 1-7 2-1. Fuse Cov

Commenti su questo manuale

Nessun commento